Product: |
Micro Direct alpha |
|
Model Number: |
MD6500 |
|
Size: |
12 x 10 x 3 |
|
Weight: |
4 pounds 7 ounces net |
|
Parameters Measured: |
FVC, VC, FEV1, FEV1 ratio, PEF, F25-75, FEF25, FEF50, FEF75, FIF25, FIF50 and FIV75 |
|
Predicted Normals: |
NHANES III |
|
Operating Temperature Range: |
ATS/ERS Limits: 63-98 F
Design Limits: 50 - 104 F |
|
Voltage/Frequency: |
110 - 250V ~ 50/60Hz |
|
Flow Detection Principle: |
Fleisch type pneumotachograph |
|
Volume Detection: |
Flow integration sampling @ 100Hz |
|
Volume Accuracy: |
Better than +/- 3% |
|
Maximum Display Volume: |
10 L |
|
Maximum Test Duration: |
90 seconds |
|
Flow Impedance: |
<0.1kPa/L/S @ 14L/s
Complies with ATS/ERS 2005 |
|
Accuracy when Operated in Operating Temperature Range Conditions: |
Flow +/- 10%
Max flow rate +/- 16L/s
Min flow rate +/- 0.02L/s |
|
Safety Standards: |
EN ISO 60601-1:2005 -1 -2 -4 |
|
Medical Safety Standard: |
Medical Devices Directive
93/42/EEC1993 |
|
Designed & Manufactured Under: |
ISO 13485:2003, FDA 21 CFR820,
CMDR & JPAL |
|
Communications: |
USB and Compact Flash |
|
Performance Standards: |
ATS/ERS 2005 & EN 13826 |
|
Printer: |
Thermal | |